Shodex Gel Filtration Chromatography (GFC) Columns |
|
GFC, using primarily aqueous solvents, is used primarily for the analysis of biological polymers, such as proteins and peptides, polysaccharides and polynucleotides.
The Support material can be silica or a polymer. Silica has the advantage that it has a larger pore volume and narrower pore size distribution. hence it generally gives sharper peaks and can give better resolution in some cases. However, large pore size silica particles are brittle and quite fragile, and if the pressure rises too high the particles collapse. The other problem as always is the pH limitation. Because silica dissolves at an unacceptable rate ove pH 7-8, it is necessary not to exceed pH 7.5 in eluents used with silica columns. In general silica has been found to give the best results for small proteins, from 10,000-1,000,000MWt, but for other samples, we recommend using a polymeric support.
pH control in GFC. The need to control the pH is really only to keep samples in solution, since the extent of ionisation does not affect retention time. However do see the note below about the addition of salt to GFC eluents. If Cl ions are present in the eluent, the eluent pH should be maintained in excess of 6.0, and if silica columns are used, the eluent should not exceed pH7.5.
Separation occurs on the basis of decreasing molecular size, which to a first approximation equates to decreasing molecular weight. No chemical interaction must occur between the sample and the packing material for a pure GFC separation to occur. However there are columns available which do offer a measure of interaction with the column in addition to the GPC separation.
Mixed mode columns. Sometimes it is hard to achieve resolution by GFC alone. In this situation, Shodex offer a range of Mixed mode columns, which allow some elements of traditional HPLC interactions to cause delayed retention of certain species on the column. In this case, the ability to derive molecular weight information from the retention time is sacrificed in order to achieve resolution.
A column is defined by its Exclusion Limit. Molecules bigger than this cannot enter the pores and pass unretained through the column. However columns with high exclusion limits offer very limited resolution of species with low molecular weight, so it is normal to require more than one column.
The total volume of the pores is known as the "Inner Volume" or "Pore Volume" of the system, Vi. The largest molecular weight components (at or above the Exclusion Limit) all elute in the Void volume of the system (V0). All components smaller than the exclusion limit of the column will elute between V0 and V0+Vi. So it follows that the bigger the pore volume of the system, the better resolution can be achieved, and this is why GFC columns are usually wider and longer than normal HPLC Columns.
Use of NaCl or NaNO3 in the eluent. GFC samples which are ionic will adopt positive or negative charges throughout the molecule. This causes parts of the molecule containing the charges to repel each other and the molecule expands significantly. Hence the apparent size of the molecule in GFC (and hence its retention time) is pH dependent. A form of ion-suppression can be used to minimise this effect, and this is achieved using approx 0.3M NaCl in the eluent. Should it be necessary to work at lower than pH6, then NaCl should not be used, and a salt such as NaNO3 used instead.
Use of organic solvents in the eluent. To avoid any hydrophobic interaction with the column, it is sometimes desirable to add polar organic solvents to the eluent, such as methanol or acetonitrile. This is generally no problem, although for some columns it is recommended not to go as far as 100% organic solvent.
Temperature control. Changes in temperature affect the viscosity of the eluent, and the kinetic energy of the samples. As a consequence, retention times are temperature dependent. For this reason it is recommended to select a temperature a few degrees above the highest ambient likely to occur in the lab, and control to this temperature accurately. However, sometimes back pressure becomes an issue, either because of the viscosity of a solvent used in the eluent (eg DMSO), or because higher salt concentrations are being used. In this situation, the use of elevated temperature (over 50oC) are recommended. However take care to note the maximum temperature which is compatible with the column chosen.
To separate a wide range of molecular sizes, there are two standard approaches:
- Use a series of 3-4 different columns connected in series, each covering a different molecular weight range. Careful selection of the columns used (eg exclusion limits 400,000 + 70,000 + 5,000) will give a fairly linear calibration curve over the desired range.
- Use 3 x Mixed bed columns. Effectively the packings from the three different column are mixed, and packed into the columns so that the samples pass through a continuous bed, and the separation occurs slowly through the columns. This gives a much better linearity and is always recommended where a suitable column is available.
Shodex have a range of columns to suit both of these approaches. Mixed bed columns are designated with an M.
Column format in GFC.
- The standard column size for GFC is 8.0 x 300mm. This is to achieve the large pore volume required, as described above. For most separations, this is the column size to choose.
- Then there is the High Resolution Format. These columns are 4.6mm x 300mm. They use a very small particle size, yet are almost as long as a standard column, so with a slower flow rate the resolution is better than either of the other two. The smaller sample size means that a smaller flow cell can be used, thus developing almost 4 x the sensitivity of a standard column system.
Guard Columns are used in GFC to prevent any chemical impurities in the polymer sample from binding chemically to the GPC column.
Sample preparation is very important for GFC, especially so for large molecules. Separation is achieved on the basis of molecular size, so it is important that the molecules are allowed to swell and stabilize in the solvent being used as eluent. This make take several hours. Where possible, the eluent should be the sample solvent. Sample concentration is also very important.
|
|
Features of columns to note are: |
Exclusion Limit |
The exclusion limit is the molecular weight at and above which the samples elute with the void volume from the column. |
Particle Size |
Typically 3-20 μm. The samller the particles the higher the efficiency (sharper the peaks) but the higher the back pressure. |
Efficiency |
Some GFC columns give sharper peaks than others. |
Solvent |
If possible, choose a column which is packed in the solventwith which it is going to be used. This minimises the effects of gel swelling caused by solvent changes, and increases the lifetime of the column. |
Dimensions |
A "standard" GFC column is 8.0mm x 30cm, but for high resolution GFC 4.6mm x 250 is available for the silica based columns. |
Temp Range |
Some solvents are viscous and need to be used at elevated temperatures. |
Max Pressure |
Some polymers can take more back pressure than others so check the maximum and do not exceed it. |
More information is available below or at the Shodex website www.shodex.com
|
| |
Column Details |
|
|
|
|
|
|
Max. Temp. (C) |
Maximum flow rate (ml/min)
|
|
|
|
150,000 - 1,000,000
Proteins
|
|
|
|
|
|
750 |
This range of columns has been designed for the analysis of small proteins, in the size range 100,000 - 1,000,000. The silica stationary phase has a hydrophilic polymer bonded on. The column can be used with up to 100% methanol or acetonitrile. Sometimes this type of column is used as the clean up stage in a column switching application, and there is a short version of the column specifically for this purpose. There is also a new range of high resolution columns, using a narrower column (4.6mmid), and a smaller particle size (3u). Espeially for small sample concentrations, this gives noticeably better sensitivity and resolution.
For pricing and ordering information click here, or for more info see www.shodex.com
|
|
|
4,000 - 500,000,000
Pullulan
|
48,000 |
|
|
|
|
|
Also known as the OHpak SB columns, these are polymer based columns for GFC. They cover a very wide molecular weight range, with an excellent mixed bed column for samples from 20,000,000 downwards. They can be used with up to 75% methanol or acetonitrile, or even 100% Dimthyl Formamide, but please check the individual column instructions, because there is some variation in this for different pore size columns.
For pricing and ordering information click here, or for more info see www.shodex.com
|
| GS |
|
3,000 - 2,000,000
Pullulan
|
|
|
|
|
|
|
The GS Series are designed as mixed mode columns, offering not only GFC separation but some degree of interaction with the column. As a consequence, for separations which cannot be resolved using GFC alone, it may well be possible to achieve resolution. A wide range of column sizes are available, but the standard size is 7.5 x 300mm.
|
GF |
|
40,000 - 10,000,000
Pullulan
|
58,000 |
2.0 - 9.0 |
5 - 9 |
60 |
1.0 |
|
The GF series of columns has been designed as multi-solvent columns for GFC/GPC. They can operate in water with up to 0.5M salt concentration, and with any concentration of methanol, acetonitrile, ethanol, THF, DMF, acetone, chloroform or ethyl acetate, and up to 50% DMSO. Standard column dimensions are 7.5 x 300mm, but many smaller diameters are available.
For pricing and ordering information click here, or for more info see www.shodex.com
|
|
|